Polyacrylamide Gel Electrophoresis
PAGE (Polyacrylamide Gel Electrophoresis), is the most widely used analytical method
to resolve separate components of a protein mixture based on size.
For this protein molecules of different shapes and sizes, need to be denatured. This is done
with the aid of SDS, so that the proteins no longer have any secondary, tertiary or
quaternary structure. The proteins covered by SDS are negatively charged. When loaded
onto a gel matrix and placed in an electric field, the proteins will migrate towards the
anode (positively charged electrode). They are then separated by a molecular sieving
effect based on size. After visualization by a protein-specific staining technique,
the size of a protein can be estimated by comparison of its migration distance with
that of a known molecular weight marker. Procedure:
1. Assembling the glass plates • Assemble the glass plates on a clean surface.
Lay the longer glass plate (the one with spacer) down first, and then place the shorter glass
plate on top of it. • Embed them into the casting frame and
clamp them properly. Make sure that the bottom ends of the glass plates are properly aligned.
• Then, place the frame on the casting stand.
2. Casting the gels
For the preparation of resolving gel, we require: dH20 4.0 ml
30% acrylamide mix 3.3 m l
1.5M Tris pH8.8 2.5 ml 10% SDS 0.1 ml
10% ammonium persulfate 0.1 ml TEMED 0.004ml
And for the Stacking gel (4.5%)
dH20 5.65 ml 30% acrylamide mix 1.65 ml
1M Tris pH8.8 2.5 ml 10% SDS 0.1 ml
10% ammonium persulfate 0.1 ml TEMED 0.004ml
1. Now, prepare the separating gel/resolving gel solution by combining all reagents except
ammonium persulfate (APS) and TEMED. 2. Add APS and TEMED to the monomer solution,
just before pouring it in between the glass slides and mix well by swirling gently. Pour
the solution until it reaches the mark on the glass slide. (It is ok if you introduce
air bubbles, add a layer of isopropanol or distilled water on top of the gel which will
level the poured gel). 3. Allow the gel to polymerize for 20-30 minutes.
4. Prepare the stacking gel monomer solution by combining all reagents except APS and TEMED.
Drain the isopropanol with strips of filter paper/ blotting paper.
5. Add APS and TEMED to the monomer solution (just before pouring it in between the slides)
and mix well by swirling gently. 6. Place a comb in the stacking gel sandwich.
Allow it to polymerize for 10 minutes.
3. Preparation of samples
Mix your protein 4:1 with the sample buffer and heat your sample at 37°C for 30 minutes
in a water bath.
4. Running the gel
1. To assemble, take out the gels from the casting frame and clamp them in the gel apparatus.(
Make sure that the short plate always faces inside and if you have only one gel to run,
use the dummy plate that is available to balance). 2. Secure the plates ,place them in the cassette
and lock it. 3. Now place the cassette in the tank.
4. Fill the inner chamber of the tank with 10x SDS running buffer (pH ~8.3).
5. Remove the comb CAREFULLY (without breaking the well).
[Now the gel is ready to load the samples] 6. Rinse the loading tip a few times with
buffer.(Make sure that all the buffer is pippeted out before loading the samples).
7. Insert the loading tip to a few mm from the well bottom and deliver the samples into
the well. Rinse the syringe a few times with buffer after loading.
8. Attach the power supply by closing the lid (Make sure that the connection is correct,
that is, (black to black and red to red). Set the voltage up to 180 V and run for 1
hour. (Don’t allow the dye front to go out of the gel).
5. Staining the gel 1. After running the apparatus, switch off
the power supply and take out the gel plates, and remove the gel.
2. Place the gel in the staining solution for 30 minutes.
3. Destain the gel until the bands are properly seen.
4. Approximate the molecular weight of the visualised protein bands by comparing them
with the molecular weight markers.
Caution:
1. Wear gloves at all times while performing SDS-PAGE.
2. To ensure proper alignment and casting, the glass plates, spacers, combs and casting
stand gaskets must be clean and dry. 3. Pay special attention while using acrylamide
as it is a neurotoxin.